Customization: | Available |
---|---|
Failure Alarm Indication: | Yes |
Application: | Two Control Modes Infant Incubator |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Brief Introduction:
Two control modes: Air mode and baby mode controlled by micro-computer;
Set temperature and air temperature are displayed separately;
Natural air flow humidity
Inclination of bassinet is adjustable
Failure alarm indication
RS-232 connector
Oxygen input interface
HB-YP90AB is configured with fixed Neonate Bilirubin Phototherapy Equipment model:HBXHZ-90(with blue fluorescence lamp)
DC motor
Standard configuartion:
HB-YP90A: Main body, cabinet, temperature controller, baby compartment, LV.pole, tray.
HB-YP90AB: Based on HB-YP90A with Neonate bilirubin phototherapy equipment model:HBXHZ-90(fixed model).
Optional Configuration:
> 37°C temperature override mode function, vertical height adjustment stand(VHA stand),
Weighing system, SPO2, Neonate bilirubin phototherapy equipment model: HBXHZ-90L (fixed model), Disposable skin temperature sensor.
Specification:
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 850VA
Mode of control: Air mode control and baby mode controlled by micro-computer
Control range of air temperature: 25ºC -37ºC(override mode: 37-39ºC is optional)Control range of baby temperature: 34-37°C(override mode: 37-38ºC is optional)Accuracy of skin temperature sensor: ±0.2°C
The temperature difference between incubator temperature and average incubator temperature(under steady temperature condition): ≤ 0.5C
Uniformity of temperature: ≤ 0.8ºC(mattress is on the horizontal position)Inclination of the bassinet: ±12°
Noise inside hood: ≤ 45dB(A)
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on.
The average value of total irradiance for bilirubin Ebi max of phototherapy equipment:
≥ 640μW/cm 2 ( XHZ-90 )
≥1300μW/cm 2( HBXHZ-90L )
Built-in timer range: 0-9999 hrs 59min
Accuracy of weight display: ±1% (equipped weighing system)
Environmental temperature:
Operating range: 20 °C -30°C (Set temperature should be higher 3°C than the environment temperature)
Ambient air movement rate: <0.3m/s
Product Packaging:
Package: each unit is packed into one carton
Dimension:
(Main body):1085mm x 640mm x 910mm
(Cabinet): 1125mm x 665mm x 725mm
Gross Weight:
(Main body):63Kg(for HB-YP90A)
75Kg (for HB-YP90AB)
(Cabinet):68Kg