Customization: | Available |
---|---|
Type: | Breathe Resuscitation Bag |
Function: | Breathe |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
EB8111:for adult and child
EB8121:for child and infant
EB8131:for infant and neonate
Optional accessories:
silicone face mask,oxygen tube,reservoir bag,
PEEP valve,B valve,airways,nose clip,mouth opener
Optional package:
PP case/paper box/poly bag
Feature
Intended Usage:
This product is mainly used for cardiopulmonary resuscitation and general artificial respiratory assistance, so that the patient can maintain normal breathing shortly before recovery.
Product Description:
- Standard accessories are with air bag, silicone mask, oxygen tube;
- Accessories such as mouthpieces, oropharyngeal airways, and lifting straps can be freely selected;
- 100% medical grade silicone, latex-free; silicone has good elasticity, high transparency, and can be reused at high temperature;
- Set 60 cmH2O safety pressure limiting valve;
- Integral intake valve design, direct connection of air storage bag and oxygen tube;
- In addition to oxygen tubes, air bags, and optional mouthpieces, oropharyngeal airway accessories, the ball, mask, and PC attachments can be sterilized by autoclaving at 134°C and reused.
- Use by a single patient;
- The non-slip effect of the rough surface of the recovery balloon can give the rescuer good operability.
|
Infant |
Pediatric |
Adult |
|||||
specifications |
infant≤10kg |
10
|
Adult>33kgs |
|||||
Min volume |
infant≤100ml |
>300ml |
>600ml |
|||||
Pressure limiting valve pressure |
40cmH2O |
40cmH2O |
60cmH2O |
|||||
Atmos valve joint for patient side |
15mm&22mm |
|||||||
Respiratory positive pressure valve |
Joint 30mm, pressure value 5-20cm H2O adjustable |
|||||||
Sacculus volume |
340ml |
680ml |
1630ml |
|||||
Gas storage bag volume |
1600ml |
1600ml |
2000ml |
|||||
Package |
56*34*46cm/8pcs |
56*34*46cm/8pcs |
56*34*46cm/8pcs |
|||||
Oxygen concentration reference |
||||||||
Oxygen flow L/min |
3 |
5 |
10 |
10 |
4 |
|||
ventilation frequency Times/min |
12 |
12 |
12 |
20 |
30 |
|||
Ventilation volumn ml |
500 |
500 |
500 |
250 |
40 |
|||
Oxygen concentration % |
60 |
86 |
95 |
95 |
92 |