| Customization: | Available |
|---|---|
| Type: | Non-Rechargeable |
| Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Features:
Three-step defibrillation process
Two-button operation
Extensive voice and visual prompts for the operator
Biphasic energy output
| Category | Specification |
| Dimensions | 303 x 216 x 89 mm |
| Weight | 2.0 kg |
| Operating Temperature | 0ºC to 40ºC |
| Operating Humidity | Relative humidity between 30% and 95% (non-condensing) |
| Storage Temperature(without battery) |
-20ºC to 55ºC |
| Storage Humidity(without battery) |
Up to 93% (non-condensing) |
| Power | Unit operates on internal battery only |
| External Electrical Connections | No external devices are attached to the unit |
| Risk Current Category | Internally powered equipment with defibrillator-proof BF type patient applied part (as per definition of IEC 60601-1 standard) |
| Biphasic Truncated exponential | Less than 30 seconds |
| 150, 150, 200J | Less than 35 seconds |
| 10 sec. to 150J | Less than 25 seconds |
| 15 sec. to 200J | 22 audible prompts |
| 9 sec. | LED prompts Two buttons - On/Off, Shock ±15% into any impedance from 25 to 175Ω .20Ω to 200Ω |
| Output Energy(J) |
Patient Impedance(Ω) |
I p1 (Amps) |
I p2 (Amps) |
Phase1 (ms) |
Phase2 (ms) |
Interval (ms) |
| 200 | 25 | 53.6 | 36.0 | 5.5 | 3.3 | 0.7 |
| 50 | 27.8 | 17.5 | 10.9 | 6.6 | 0.7 | |
| 75 | 18.2 | 12.0 | 16.3 | 8.9 | 0.7 | |
| 100 | 13.4 | 9.0 | 19.7 | 9.6 | 0.7 | |
| 125 | 10.5 | 7.2 | 20.5 | 11.2 | 0.7 | |
| 150 | 8.6 | 6.0 | 21.6 | 10.4 | 0.7 | |
| 175 | 8.2 | 6.1 | 21.8 | 11.2 | 0.7 |
| Non-rechargeable | 12V d.c. 2.8Ah |
| Capacity | 100 discharges at 200 Joules or 120 discharges at 150 Joules |
| Shelf Life (25ºC±15ºC) | 5 years |