Customization: | Available |
---|---|
Type: | Infant Care Equipment |
Function: | Infant Care |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Item | specifications |
heating voltage | ≤300W±10% |
Incubator temperature control range | 25-37ºC |
Skin temperature control range | 34-37.5ºC |
Skin temperature sensor accuracy | ≤0.3ºC |
Bed surface temperature average | ≤1.5ºC |
Humidity accuracy | ≤10% |
Humidity display range | 30-99% |
Temperature fluctuation | ≤0.5ºC |
Temperature rise time | <40min |
Temperature deviation alarm range | For incubator: ±3ºC |
For skin: ±1ºC | |
indoor noise of incubator | ≤45dB |
Alarm noise | Indoor: ≤70dB |
Outdoor: ≥60dB | |
overtemperature alarm | ≤38ºC |
Second overtemperature alarm | ≤40ºC |
Alarm of sensor fault | response time <4s, Continuous alarm |
Power failure alarm | response time <5s, Continuous alarm |
Air blower stopping alarm, air-duct block alarm | response time <30s, Continuous alarm |
Lack water alarm | Inlet tube water level is under the sign of "MIN" |
Upper-body overturn angle | 30ºC |
Electric lifting distance | 150mm |
Inclination of large bassinet | 0-8°stepless |
Temperature display mode | LED digital display |
Infant incubator body material | Organic glass, double walls hood, cabinet case |
Oxygen concentration measurement(optional) | 0~100% |
Panel automatically locked | yes |
Operation window | No less than 5 pcs, with auxiliary air sensor |
Panel feature | Touch panel, water-proof, sensitivity and reliability |
Low battery alarm | <4.5V |
Safety performance | According with class I, IEC601-1, 1998.GB9706.1-1995 and GB11243-2000 |