Customization: | Available |
---|---|
Type: | Anaesthesia Machine |
Certification: | CE, FDA, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Technical Specifications | |
Physical Specifications | |
Dimensions: | 70*85*140cm |
Weight: | 65kg |
Casters: | Front wheels with lock |
Screen: | 12.1 inch LCD |
Working Specifications | |
Power | AC 220V±10%, 50Hz±2% |
Pipeline Supply | |
Gas configuration | O2, N2O, Air |
Pipeline input range | 280-600kPa |
Battery Power | |
Run-time | 2180 mins |
Pre-setting Functions | |
Language | English |
Calibration | Automatic |
Applications | |
Patient range | Audit/ Pediatric |
Ventilator Specifications | |
Modes of Ventilation | |
vcv | Volume Controlled Ventilation |
SIMV-V | Synchronized Intermittent Mandatory Ventilation-Volume |
SIMV-P | Synchronized Intermittent Mandatory Ventilation-Pressure |
MANUAL | Manual |
A/C | Assist /control Ventilation |
PCV | Pressure Control Ventilation |
PSV | Pressure Support Ventilation |
STAND-BY | STAND-BY |
Ventilator Parameter | |
Working type | Electronic control, pneumatic driven |
Tidal Volume range | 20-1500ml |
Pressure range | 5-60 cmHaO |
Rispriation rate range | 1-100 bpm (1-40bpm under SIMV) |
l:E range | 4:1-1:8 |
Ptrigger range | -20-20 cmHzO (based on PEEP) |
Flow trigger | 0-15L |
Minute volume range | >18L/Min |
SIGH range | 0-5/100 |
System Standard | |
Hypoxic guard system | N2O cut-off valve, O2 concentration >25% |
Safety valve | <12.5kPa |
Fresh gas compensation | 25-75L/min |
Volume of CO2 absorber | 2 L |
Flow meter | 5 tubes, Cascade 0.1-15L/min O2 |
0.1-15L/min N2O, 0.1-15L/min air | |
Vaporizer | 2 Selectetac, maximum 2 mounts(lsoflurane/Enflurane/Sevoflurane/Halothane) |
Monitoring | |
Display graphics | Waveforms of P-T, F-T, V-T, loops of P-V,V-F,F-P(Zoom in/out, comparison, freeze) |
Types | VT, MV, BPM, Paw, Lung compliance, Inspiratory Platform,F1O2 |
Alarm Limits | |
Audio/Visual | No tidal volume, MV,Paw, FiO2 limitation O2, |
Air failure, AC power failure, Battery low | |
PEEP | |
Type | Integrated, Electronic control |
Range | OFF, 3-30cmH2O |
Optional | |
1.AGSS | Anesthetic Gas Scavenging Systems |
2.Suction Unit | |
3.Communication Interface | Ethernet, USB, VGA |
4.Auxiliary Oxygen meter |