Hbhkn-2001 Heatable and Adjustable Neonate Resuscitation Radiant Warmer

Type: Infant Incubator
Function: Baby Care
Theory: Infant Incubator
Certification: CE, ISO13485
LCD Display: With LCD Display
Group: Children
Customization:
Diamond Member Since 2013

Suppliers with verified business licenses

Manufacturer/Factory

Basic Info.

Model NO.
Advanced infant resuscitation radiant warmer
Three Control Modes
Pre-Warm,Manual,Baby Mode Control
LCD Display Screen
Stored and Indicate Information
Failure Alarm Indication
Yes
The Angle of Warmer Module
Adjustable Horizontally
Inclination of The Bassinet
Trendelenburg 15and Reverse Trendelenburg 20
The Panels Around Bassinet
Turn Outward
Mattress
Heatable and Controllable
LED Phototherapy Function
Yes
Transport Package
Export Wooden Package
Specification
ISO, CE
Trademark
WANROOEMED
Origin
Jiangsu, China (Mainland)
HS Code
94029000
Production Capacity
500set/Sets Per Month

Product Description

Hbhkn-2001 Heatable and Adjustable Neonate Resuscitation Radiant Warmer
Hbhkn-2001 Heatable and Adjustable Neonate Resuscitation Radiant Warmer
Brief Introduction:
Three control modes: Pre-warm mode control, manual mode control and baby mode control by micro-computer;
The information can be stored, consulted and indicated by the LCD display screen;
Failure alarm indication;
The angle of warmer module is adjustable horizontally;
Inclination of the bassinet is adjustable;
The panels around bassinet can be turned outward;
Mattress is heatable and temperature is controllable;
With LED phototherapy function;
APGAR timer;
Vertical height adjustment (VHA) stand.
 

Standard Configuration:
Main body (including the warmer module, controller, bassinet, main column and VHA stand), skin temperature sensor, J.V. Pole, mattress, tray and panels
 

Specification:
Power requirement: AC22OV-230V/5OHz or AC11O-120V/50-6OHz, 900VA Mode of control: Pre-warm mode, manual mode and baby mode are controlled by micro - computer
Baby mode temperature control range: 34.5-37.5°C
Accuracy of skin temperature sensor: ±0.2°C
Temperature uniformity of mattress: <= 2°C
Mattress heating control range: 25-38°C
Accuracy of mattress temperature display: ±0.7°C
Angle of warmer module: ±60°
Inclination of bassinet: The Max. angle on trendelenburg is 15°, on reverse trendelenburg is 20°
APGAR timer: audible and visible alarm when the device runs to 50"~ 1', 4'50" ~5', 9' 50"~ 10'
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on.
Max. bilirubin radiation on the mattress effective range: 1mW/cm2
Average value general bilirubin radiation of on the mattress effective range:
≥0.8mW/cm2
Bilirubin uniformity of radiating on the mattress effective range: >0.4
 

Environmental temperature:
Operating range: 18 °C -30°C
Ambient air movement rate: <O.3m/s
 

Product Packaging:
Package: each unit is packed into one carton
Dimension: l500mmx8l0mmxl475mm
Gross Weight: 250 Kg

 
Details:
Hbhkn-2001 Heatable and Adjustable Neonate Resuscitation Radiant Warmer
Company Information:
Hbhkn-2001 Heatable and Adjustable Neonate Resuscitation Radiant Warmer

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Diamond Member Since 2013

Suppliers with verified business licenses

Manufacturer/Factory
Management System Certification
ISO 9001, ISO 13485
Import & Export Mode
Have Own Export License