Item |
specifications |
heating voltage |
≤300W±10% |
Incubator temperature control range |
25-37ºC |
Skin temperature control range |
34-37.5ºC |
Skin temperature sensor accuracy |
≤0.3ºC |
Bed surface temperature average |
≤1.5ºC |
Humidity accuracy |
≤10% |
Humidity display range |
30-99% |
Temperature fluctuation |
≤0.5ºC |
Temperature rise time |
<40min |
Temperature deviation alarm range |
For incubator: ±3ºC |
For skin: ±1ºC |
indoor noise of incubator |
≤45dB |
Alarm noise |
Indoor: ≤70dB |
Outdoor: ≥60dB |
overtemperature alarm |
≤38ºC |
Second overtemperature alarm |
≤40ºC |
Alarm of sensor fault |
response time <4s, Continuous alarm |
Power failure alarm |
response time <5s, Continuous alarm |
Air blower stopping alarm, air-duct block alarm |
response time <30s, Continuous alarm |
Lack water alarm |
Inlet tube water level is under the sign of "MIN" |
Upper-body overturn angle |
30ºC |
Electric lifting distance |
150mm |
Inclination of large bassinet |
0-8°stepless |
Temperature display mode |
LED digital display |
Infant incubator body material |
Organic glass, double walls hood, cabinet case |
Oxygen concentration measurement(optional) |
0~100% |
Panel automatically locked |
yes |
Operation window |
No less than 5 pcs, with auxiliary air sensor |
Panel feature |
Touch panel, water-proof, sensitivity and reliability |
Low battery alarm |
<4.5V |
Safety performance |
According with class I, IEC601-1, 1998.GB9706.1-1995 and GB11243-2000 |